Question and Answers The mobile unit shall be equipped with an Image Analysis Station for image registration quality assurance" is this in addition to the operator's console? No reading station is req...
Question and Answers The mobile unit shall be equipped with an Image Analysis Station for image registration quality assurance" is this in addition to the operator's console? No reading station is required beyond operator s console. Is a Medrad Spectris Injector acceptable, or does it need to be a Medrad Spectris Solaris EP? Solaris injector is required Contrast warmer is not required. Do you need application training beyond the 1 day we provide at no additional cost. We do not have an Invivo/ Foot /Ankle Coil, but the unit does have a Quad Foot/Knee Extremity Coil. Additionally, the unit has a breast coil and a wrist coil. Techs currently use foot/ankle boot coil. Hard drives will need to be removed after the lease Yes this is a SDVOSB set-aside