Statement Of Work MUST be Compatible with Phillips vital sign monitoring equipment currently used at the facility Due to periodic saline shortages we require a resevour Water Temperature Control setti...
Statement Of Work MUST be Compatible with Phillips vital sign monitoring equipment currently used at the facility Due to periodic saline shortages we require a resevour Water Temperature Control setting range: 39.2-104.0 F Control accuracy: + or 0.3 C Display measurement accuracy: + or 0.2 C Display/resolution setting: 0.1C Default setting: 104.F (40.0C) Patient Temperature Control setting range: 89.6-100.4F Control accuracy: = or 0.1 C Default setting: 98.6F Circulating Fluid: Sterile distilled water or distilled water Thermal transfer devices such as blankets, vests, and leg wraps Patient temperature probes Reusable adapter cables Reusable hose sets Reusable patient temperature output cable Must be capable of supplying either an individual or multiple thermal transfer devices simultaneously with those circuits being monitored separately. Must have 3 operating modes: Automatic, Manual, and Monitor Closed loop control Controller alarms visual and audible indications for when safety parameters are exceeded or it detects system function or performance irregularities. A display to inform of system status or when the user must confirm a setting selection.