Electrosurgical unit (ESU)
Contract seeks newly manufactured, unused electrosurgical equipment and compatible accessories, not used, refurbished, reconditioned, remanufactured, or demonstration equipment, all in full working condition at delivery. The equipment must include a touchscreen interface, tissue impedance monitoring at a minimum rate of 434,000 times per second, output frequency of 400–500 kHz, and support monopolar CUT (Pure/Blend), monopolar COAG (Fulgurate/Spray/Soft), bipolar coagulation, and vessel sealing. The contractor must provide all accessories needed to operate every mode, including a switched monopolar electrode and holster, patient return electrode, bipolar forceps with cord and footswitch, and a switched vessel sealer. All accessories must be compatible with the delivered equipment and ready for use upon delivery. Compliance with applicable Federal, state, and local safety standards is required.