Inactive
Partial Small Business Set-Aside (FAR 19.5)
Notice ID:RFQ323763_BLN
3 EA PORTABLE PARTICLE COUNTER Technical specification for a cleanroom particle counter Flow rate updated to 100 L/m. Portable particle counter compatible to iso class 4 cleanroom or better. Multi-cha...
3 EA PORTABLE PARTICLE COUNTER Technical specification for a cleanroom particle counter Flow rate updated to 100 L/m. Portable particle counter compatible to iso class 4 cleanroom or better. Multi-channel detections of particles with the most sensitive channel to be at or better than 0.3 mm. Typical particle sizes are 0.3, 0.5, 1.0 and 5.0 mm. Sample flow rate one SCFM (standard cubic foot per minute) or higher, adjustable. Resolution: Exceeds requirements of ISO 21501-4 of 15%. Maximum False Count: 2.7 / CM ? 0.3?m (95% UCL) per ISO21501-4 HEPA filtered internal exhaust, meets, or exceeds ISO class 3. Tabletop mountable sample probe stand per unit. Sample tubing 3-6 FT per unit. One zero count filter per unit. Internal memory for sample data retention. Wall Power: Universal 100-240VAC Input. Internal battery to allow 4 hours of continuous sampling. Connectivity includes RS-232, ethernet, WIFI, and USB. Certificate of calibration to ISO 21501-4 at an ISO 17025 accredited laboratory. Availability of both on-site and off-site calibration services. Compliant with FDA 21 CFR Part 11 data integrity standards. NRTL CERTIFIED