LS-CAT 21-ID beamline requires high precision X-ray microdiffractometers for two of their end stations, 21-ID-F amd 21-ID-G.. These end stations are currently operating with Maatel MD2 microdiffractom...
LS-CAT 21-ID beamline requires high precision X-ray microdiffractometers for two of their end stations, 21-ID-F amd 21-ID-G.. These end stations are currently operating with Maatel MD2 microdiffractometers acquired in 2005. These MD2 microdiffractometers are end-of-life and do not have the accuracy to center micron sized crystals in the post APS-U X-ray beam. The vertical rotation axis of the MD3-UP is essential for reducing the sphear of confusion to an acceptable level. Nanometer sample alignment Kappa axis with <200nm sphear of confusion Dynamic accuracy <1mdeg at 10deg/s On-axis microscope for visualization and centering Schedule - The two micro-diffractometers shall be delivered within nine (9) months of the award of the contract.