Inactive
Notice ID:W81K02-20-Q-0142
This requirement is for a lease of a hematology analyzer (including reagents, supplies, training, and service of the system). System MUST be authorized and compatable to connect with Roche Diagnostic'...
This requirement is for a lease of a hematology analyzer (including reagents, supplies, training, and service of the system). System MUST be authorized and compatable to connect with Roche Diagnostic's Total Laboratory Automation sytem. -Must have a conditional ATO at a minimum. -Must be able to provide fluorescent dye based parameters to include immature platelet fraction, blast cells, abnormal cells and retic hemoglobin equivalent. -Must be able to provide an analyzer to perform Erythrocyte Sedimentation Rate (ESR) testing. -Must perform a six part differential to include immature cells absolute and percent. -Must provide a scattergram the separates lymphocytes. -Must be able to correct for interfering substances such as NRBC, platelet aggregation, and etc. -Hemoglobin must be determined by direct spectrophotometric analysis. -Capable of performing automated cell count on body fluids.