Inactive
Total Small Business Set-Aside (FAR 19.5)
Notice ID:N0020322STRLCABINET
Requesting 6 cabinets with Exterior dimensions should be 30.5" W x 32" D x 77.5" H, Shelf quantity and dimensions: 5 - 27.75" W x 18" D, Overall Payload dimensions: 28" W x 19" D x 42" H, and estimate...
Requesting 6 cabinets with Exterior dimensions should be 30.5" W x 32" D x 77.5" H, Shelf quantity and dimensions: 5 - 27.75" W x 18" D, Overall Payload dimensions: 28" W x 19" D x 42" H, and estimated weight: 450lbs. Requesting installation and training for all 6 cabinets. Cabinets should have the following characteristics: - Exceeds requirements of VA Directive 1116(2); CSA (Canadian Standards Association) Z314.15-10; ANSI/AAMI ST79-2006; ANSI/ASHRAE/ASHE Standard 170-2017. Conforms with UL60601-1 - Have an onboard custom control system that provides easy operation utilizing a digital display for monitoring temperature, humidity, pressure and status. - Maintains a temperature of 66F to 72F - Contain integral pressure monitoring with alarm - Maintains Humidity at 20% to 60% RH - Must have steam humidification utilizing standard demineralized water - Maintains positive pressure +0.03"WC - Should have an integral 3/4 gallon reservoir filled through front spout access - Meets IEC 60601-1 Medical Standard for Power Supplies - Requires a 120V hospital grade 10ft long plug-in-cord - 55-59 DBA at 3 feet (1 meter) - Requires a HEPA Filtration designed to exceed ISO 6 Cleanliness or better in some environments - Optional ducted heat exhaust - Requires to have 230 air changes per hour - Requires UV-C light (bathes HEPA filter) - Equipped with a Integrated condensate control system - Capability to interface with VAnetwork environmental monitoring system via an access port.