See Attachment Responses to questions sent during sources sought: Modular pressure controller 1. Requirement #2 states a range from 6 kPa to 100 kPa but also states that the full scale is 350 kPa. Can...
See Attachment Responses to questions sent during sources sought: Modular pressure controller 1. Requirement #2 states a range from 6 kPa to 100 kPa but also states that the full scale is 350 kPa. Can you please clarify the requirement for this line item ? - The overall range of the item is 7MPa (~1015 PSI). o This range is broken up into 3 separate transducers. ยง Transducer 2: Range of 6 kPa to 700 kPa with an accuracy of +/-(0.007% F.S. {"Full Scale" = ~700 kPa/~100 PSI}). 2. We strongly suggest that you include our contamination prevention system (CPS) accessory which protects the 6270A from damage if the TI has residual contaminants that can migrate into the 6270A during operation. Please share the following link with the end user and let us know if you want to modify the solicitation. 6270A Contamination Prevention System Video | Fluke (flukecal.com). - We only use this on clean, dry nitrogen. We do not need to purchase an additional Contamination Prevention System. Thank you though!