Inactive
Notice ID:FDA-RFQ-20-1231451
The Contractor shall provide a Steritest Symbio ISL Pump or equivalent. The instrument shall meet or exceed the following requirements: The instrument shall consist of a peristaltic pump compatible wi...
The Contractor shall provide a Steritest Symbio ISL Pump or equivalent. The instrument shall meet or exceed the following requirements: The instrument shall consist of a peristaltic pump compatible with Steritest canister devices. The pump shall have a table integrated design to be installed into a circular cut-out, measuring 210mm, on a horizontal worksurface of a Getinge Isoflex-S 3-Glove Isolator, serial number #3089. The pump head must be capable of automatic opening and closing. The pump housing is to be constructed of 316L stainless steel. The pump housing and components to be resistant to vaporized hydrogen peroxide gas. The peristaltic pump must have an adjustable speed with a maximum speed of at least 200 rpm. Pump shall be equipped with a keypad for selecting instrument settings. Pump shall be equipped with a timer function. Pump shall be equipped with a pressure control system. Pump shall be equipped with a color LCD user interface screen. Pump shall have the ability to store at least 200 filtration protocols. The pump body must not exceed the following measurements: 10 inches in width, 13 inches in depth and 8 inches in height.